1 The Safe Use of Unit
1.1 Personal Safety and Safeguards
Make sure that the power source has dependable ground wire! If it is not connected to the ground, the casualty accident may be caused.
Please do not use the instrument in dangerous environment! The shell is not complete airtight, it is possible that the ingress of sparkle and gas can cause explosion.
When using chemicals or solvent, you should follow the instruction manual of manufacturer and safety specifications of laboratory!
All of the Karl Fischer reagents are flammable and toxic. If skin is inadvertently in touch with Karl Fischer reagents, you should immediately wash skin with plenty of water. If eyes are inadvertently in touch with Karl Fischer reagents, you should immediately wash skin with plenty of water and consult doctors.
1.2 The Safeguards for Operation
2. Application Scope and Principle of Measurement
2.1 Measurement of Moisture
The Karl Fischer method, a classical and the most dependable method, is used to measure the trace moisture of different substance. Karl Fischer coulometric method is more suitable for measuring the substance, the absolute water content of which is less than 100μg. The Karl Fischer coulometric method is applied in this unit. This unit is especially suitable for measuring the trace moisture in the sampling, the trace moisture in sampling can even be precisely measured if adopting the proper operation.
2.2 The Brief Characteristics of Unit
Taking the singlechip control as the hardware basis, which is developed through integrating with comprehension of physical and chemical phenomena in the measurement process, The analysis speed is swift, the precision is high, and the measurement results are directly displayed on the LCD, many analysis methods can be chosen freely to set , the record can be printed, the operation time can be set, in conclusion, this unit is a high-efficient, fully automatic delicate analysis instrument.
2.3 Application Domain
For the fields, such as petroleum, the petrochemical industry, electric power, pharmaceutical, pesticide, environmental protection, geology, and foodstuff, you may use this unit if there is a need of analyzing the water content in raw material, intermediate and product. For example, some state standards, such as GB/T 3727-2003 “Measurement of Moisture in Industry-used ethene and Propene” and GB/T 7600-1987 “Measurement Method of Water Content of Moisture in Transformer Oil (Coulometric Method)” have definitely specified that the Karl Fischer coulometric method instrument must be used.
3 Technique Parameters
3.1 The adopted method: Karl Fischer coulometric method
3.2 Display: LCD with English character
3.3 The control of electroanalysis: automatically controlled by electroanalysis electric current(the maximum electric current: 400mA);
3.4 Measurement range: 3ug ~ 200mg water.
3.5 Titration mode and speed: coulometric titration, 2.4mg/min(the maximum)
3.6 Sensitivity valve: 0.1μg H2O
3.7 Accuracy degree: postulating the absolute water content of sampling is X, the relative error is
3.8 Print: heat sensitivity line printer
3.9 Power source: AC 220V, 50Hz;
3.10 Power: 35W
3.11 Temperature of use environment: 5~
3.12 Humidity of use environment: <80%;
3.13 Weight: about
3.14 Outside dimension: length--
4 Structure Characteristics and Instructions for Keys
4.1 Total line-up
Figure 1 Total Line-up of unit
1. main unit; 2.stirrer and titration cell;
4.2 Keyboard and key function
[AUTOSTART Key]: Measure the water content i when use together with the sampling and sample injection instrument of liquid hydrocarbon flash distillation and gasification;
[START Key]: pressing this key, the feeding will begin after the delay time.
[ENTER Key]: Use it to end the data input and execute the chosen order.
[MENU Key]: Use it to return to the previous menu.
[DATA Key]: Use it to inquire the historical measurement data.
[HELP Key]:
[RESET Key]: Use it to restart the instrument program, generally it can be used when the instrument exceptional.
4.3 Brief introduction of display window
1. Main menu; 2. Main interface; 3. Information prompting column; 4. time display space
Water content: the water content of sampling, unit: µg or mg;
State: the digital display of balance point. The balance point of unit is 40;
Speed: it is the electrolytic speed, unit: µg/s;
Delay time: it is the time after pressing START key until turn on the electrolytic current;
Concentration: the sampling’s concentration of water,
4.4 Rear panel of main unit
Figure 2 Rear Panel of Main Unit
1. Stirrer cable connector;
2. Single connector (the single cable for the sampling and sample injection instrument of liquid hydrocarbon flash distillation and gasification (model LG-5) is connected here. The automatic control of feeding is finished by it).
3. 232 string interface, the single cable for computer is connected here.
4. Reserve.
5. Fan.
6. Power cable connection port.
7.
4.5.The Introduction for Stirrer
Figure 3 the Installation Drawing for Stirrer
1. Stirrer;
2. Plug for main unit;
3. Detector electrode connector;
4. Electroanalysis electrode connector;
5. Generator solution cell,
6. the Base of generator solution cell,
7. Stirrer speed adjuster (the speed is adjusted to the value that can cause vortex, please do not adjust the stirring speed during the measurement process). ①the minimum stirring speed or the stop of stirring; ②this speed is suitable when the electrolytic solution is less than 150mL; ③this speed is suitable when the electrolytic solution is more than 150mL; ④the maximum speed, generally this speed is not chosen.
4.6 The Structure Introduction for Generator Solution Cell
Figure 4 the Structure of Generator solution cell
1. Generator solution cell; 2.Detector electrode; 3. Versatile hole and sealing plug;
4. Electrolysis electrode or cathode solution cell; 5.drying tube;
6. Injection port of cathode solution and plug for cathode solution cell;
7. Sample injection port and plug;
8. The upper layer is platinum net of cathode, the middle layer is porous ceramic panel, the lower layer is platinum net of anode;
9. Stirrer bar;
10. Cathode solution cell connector;
11. Detector electrode connector
4.7 Printer and set-up of printer roll paper
1. Printer cover; 2. Roll paper, 3. Paper entry, 4. Paper change handspike
5. The operational instruction of the unit
5.1 main menu instructions
Main menu contains: ‘set’、‘method’、‘func.’、‘super’ and ‘help’ . First of all , press the [left] key or [right] key while operating, select the main menu, then press the [enter] key to enter the corresponding submenu , press the [up] key or [down] key ,select the corresponding submenu , then press the [enter] key , the instrument will begin to run the function that corresponding to the submenu , if you want to return to the previous menu ,please press the [enter] key.
5.2 The setting direction of the ‘method ‘menu
Method menu contains :‘method No.1~No
Select the ‘method’ in main menu and press the [enter] key , select the method that you want to set , then press the [enter] key , entering the method you want to set , select the option that you want to set , press the [enter ] key, the setting items are ‘sample ID’、‘sample type’、‘sample unit ’、‘conc. unit’、‘delay’、 ‘Precision’ 、 ‘max. current’ 、‘mass method’.
%(v)(volume)is the volume ratio, be sure to input the volume of the sample previous to the analysis. %(v)( density ) is the density ratio, be sure to input the density previous to the analysis, if you input the mass previous to the analysis , the mass must be inputted firstly.
5.3 the setting direction of the ‘set’ menu :
‘set’ menu contains: ‘time’、‘method’、‘print mode’、‘printer’. Select the ‘set’ key , press the [enter] key, then select the corresponding item , press the [enter] key.
5.4 The setting direction of the ‘func.’ menu:
‘function’ menu contains : ‘analyse’、‘automatic’、‘print’、‘reagent’. Select ‘func.’, press the [enter] key, then select the corresponding item, press the [enter] key.
5.5 The setting direction of the [super] menu:
‘Super’ menu is our company’s debugging menu; we suggest users not to use it. ‘super’ menu contains : ‘debug’、‘Balanced’. Select the ‘super’ key, press the [enter] key, then select the corresponding item, press the [enter] key.
If you select the ‘electrolyse’ , the inputting data range is 0~360( including 0 and 360) , press the [enter] key , display interface will show the corresponding data you set , generally input 0 and 360 .
If you select the ‘detector’, entering the detector debug interface, you can carry on the detector debug work, make the detector electrode short-circuit or increase a resistance of the fixed value, display interface will show the corresponding data.
5.6 The setting direction of the ‘help’ menu:
Select the [help] key , press the [enter] key , then select ‘aboult’ or ‘ver.’ , then press the [enter] key , the interface will show the information of our company's name 、network station、firmware version and so on . Return to the previous menu, please press the [menu] key.
5.7 Attention
6 Principles of Measurement and Method of Application
6.1 Principles of Measurement
The reaction of Karl Fischer reagent and water:
I2+SO2+
The used reagent solution is composed of certainty concentration iodine, I- and carbinol dissolved with sulfur dioxide and so on. The measurement principle is that conductive capability in the balanced system of certainty concentration iodine and I-, after switching on electric current to two electrodes, the following reactions will begin:
2I--2e→I2
I2+2e→2I-
There is superfluous I- in the solution, but concentration of iodine will not cause great influence on the conductivity of solution, while the change of iodine’s concentration can cause the change of solution’s conductivity. The more the concentration of iodine is, the more the change of solution’s conductivity, we take this scope as the effective range of measurement, through measuring the current strength generated by the voltage of platinum electrodes, this current strength can reflect the conductivity of solution. This instrument converts the measured value of current strength into the digital quantity of inverse proportion, which reflects the status of solution, the more the digital quantity is, the less the conductivity is. Generally, the balanced point of iodine concentration is corresponding to 40(the digital quantity)
The water concentration of the point is lower also, if it is lower than water concentration of sample, it can be ignored. When a certain mass of water enters into the balanced system, balance constant of main reaction is big especially, reaction progresses thoroughly, so concentration of iodine is reduced, conductivity of solution is reduced, state number of solution is increased, the above is measurement principle.
6.2 Basis of Working and Metrology
According to the above-mentioned principle, we know that the preparing state of instrument is to maintain a certain concentration of iodine. Environment’s moisture is uninterrupted going into the balanced system, which causes the reduce of concentration of iodine. To maintain the concentration of iodine, we make that I- in the solution lose electron to change into iodine by adding positive electricity of electrolyte electrode, when speed of consuming electrolyte electrode is equal to speed of electrolyte generating electrolyte electrode, there is balance state in the system solution. Current, which keeps the balanced state, is called blank current, the current consumes the inbreak moisture, after conversion, the blank current is expressed by the speed of consuming inbreak moisture, namely, μg/s.
When beginning measurement, adding water is made to reduce concentration of iodine, iodine is generated in anode by electrolysis, Based on Faraday's law, all generated iodine are direct proportion to consuming electric quantity:
2I--2e→I2
amount of iodine joining in reaction is equal to amount of water. Injecting sample into electrolytic solution, water in sample joins in reaction, measuring iodine change in reaction course by instrument, consume of iodine is calculated according to electrolyzing iodine, while the amount of iodine can be calculated by the instrument, the mass of water is showed directly in LCD. Automatic control system of electrolysis current is used in the instrument, size of electrolysis current is automatic controlled according to water content of sample, the maximum is 400 milliampere. In the course of electrolysis, water content is reduced step by step, electrolysis speed is reduced with proportion till balance point, corresponding titration end point control unlock of circuit. This system insures a high precision, high delicacy and high speed in analysis process. In addition, in the course of measuring, some factors are hard to avoid, such as inbreak moisture of atmosphere, which cause blank electric current of generating in titration cell. Because there are function of recording and deduction blank electric current in the instrument, the mass showed in LCD is real mass of water in sample.
6.3 Switch on
After the preparation of generator solution cell, switch on 220V power source. Turn on switch of unit, the instrument will enter the initial interface, the buzzer will sound at the same time. There is the electrolysis current after 2 seconds, the instrument will automatically search balanced point, the status display will be close to 40.
6.4 The Automatic Deduction of Blank Electric Current
The blank electric current influences the precision of measurement greatly. This instrument has the functions of automatic-measuring and automatic deduction of blank electric current. The instrument can finish the automatic deduction of blank electric current after the 2mins running.
When measuring the water content in sample, we wish the lowest blank electric current in order to get a high precision. But the electrolytic current still shows that it is larger than 0 and steady, now, press “SAMPLE INTRODUCTION” key, after a moment, the buzzer will sound, the display of water content is still zero or smaller than a numerical value(this numerical value should be smaller than or equal to the 10% of water content of sample. Now, the measurement can be implemented.
When measuring the trace moisture(less than 10 ug), the generator solution cell should be desiccated thoroughly, when the status value is smaller than 45, the measurement can start.
6.5 The Cleaning, Drying and Assembling of Titration Cell
These processes should be operated outside the stirrer apparatus. The following are the concrete steps:
6.6 How to Reduce Blank Electric Current
If the status display is larger than 70, speed is larger than 0.5μg/s, these indicate that the generator solution cell is not desiccated thoroughly. The reason is that there is moisture on the inner wall of titration cell. You should switch off the power, take off the titration cell, lean it slowly and sway it, to make the moisture into electrolytic solution, then set titration cell again, switch on, continue titration, this step can be repeated several times. Through the above operations, blank electric current will lower to the minimum value. If the blank electric current still does not lower, this may be caused by the intake moisture from atmosphere or the ceramic filter panel absorbs moisture. Now, do the following inspections:
(1) whether the sealing of titration cell’s abrasive mouth is good, whether the smearing of grease is even;
(2) whether the silica gel is effective, whether there is reagent in drying tube; if the silica gel has become to light blue or red, replace it immediately;
(3) whether the cleaning and drying of cathode solution is good;
(4) whether the sample injection port is hermetic, whether the pinhole of silica gel packing is leak;
(5) whether the liquid level of cathode is parellel or higher than the liquid level of anode;
(6) whether the compounds contain aldehyde or carbonyl group is poured;
6.7 The Common Standardization of Instrument
When the instrument reached the balanced point and is steady, you can standardize it with pure water. The standardization mode is as follows:
(1) Ingest 0.1 microlitre distilled water with 0.5 microlitre sample introduction instrument, prepare for the sample injection;
(2) Press “SAMPLE INTRODUCTION” key to start delay time;
(3) Before the end of delay time, pour the distilled water into the anode cell through plug for sample injection port, the pinpoint should be inserted into the electrolytic solution, and avoid contacting with inner wall of titration cell or electrode, after the end of delay time, the titration will automatically start;
(4) When the buzzer sounds, the instrument reaches the end point, the display result should be 100±10microgramme water(including sample introduction error), generally standardize it 2~3 times, when the result is within the error range, the measurement can begin.
6.8 The Special Standardization Method
When the instrument is used for a long time or there is a need of high precision, you can implement the special standardization. The special standardization should be in accordance with the state standards, the method is benzene—saturated water standardization.
In order to inspect the coulometric moisture meter, you need to make the saturated solution of benzene, get water content through accurate measurement of liquid temperature. The standard sample of benzene is known, you can inspect the inspection limit, repeat error and basic error of instrument through the control of sample injection quantity.
a) the coulometric moisture meter to be inspected
b) analytical balance(precision is 0.1mg, it must has certificate for inspection of metrology);
c) small syringe(0.1mL,1mL,5mL);
d) wide-mouth bottle(250mL) with 2 holes in the plug, one is sample injection hole, the other is the hole of thermometer;
e) thermometer(0~
f) Karl Fischer reagents(made by Daqing Rshang Instrument Co. Ltd);
g) Benzene(analytically pure)
a) the forming of standard sample of benzene: pour 120mL distilled water into the wide-mouth bottle, then pour 60mL~70mL benzene into it, wait 1 hour to arrive the saturation status, wait for measurement;
b) Run the instrument, implement the measurement after the instrument stabilizes;
c) Clean the small syringe, that is, take in the standard sample of benzene, repeat it several times, the first cleaning should be more than 5 times, the cleaning times in analysis should exceed 3 times, the used liquid should be put back to reclaiming bottle, cleaning it until there is no water in the syringe;
d) Take 5μL standard sample of benzene, at the same time, record the temperature of benzene’s saturated solution(the precision is
e) According to above method, take 0.15mL, 1mL standard sample of benzene respectively to measure, you also use different volume according to the concrete cases.
a) Read the numerical value of trace moisture from display screen, the precision is 0.1μg.
b) Search the saturation water solubility of benzene relevant to the experiment temperature from Table 1 “the Data Table for Saturated Water Solubility of Benzene”, calculate the water content of benzene, namely, it is the theoretical value, the precision is 0.1μg. The record format is listed in Table 2 “the Inspection Results of Saturated Water in Benzene for Different Sample Introduction Quantity”.
Table 1 the Solubility Data for Saturated Water in Benzene
temperature
(℃) |
water content
(μg/kg) |
temperature
(℃) |
water content
(μg/kg) |
temperature
(℃) |
water content
(μg/kg) |
10 |
0.440 |
20 |
0.614 |
30 |
0.859 |
11 |
0.457 |
21 |
0.635 |
31 |
0.888 |
12 |
0.474 |
22 |
0.655 |
32 |
0.918 |
13 |
0.491 |
23 |
0.676 |
33 |
0.947 |
14 |
0.508 |
24 |
0.696 |
34 |
0.977 |
15 |
0.525 |
25 |
0.716 |
35 |
1.006 |
16 |
0.543 |
26 |
0.745 |
36 |
1.055 |
17 |
0.561 |
27 |
0.773 |
37 |
1.104 |
18 |
0.579 |
28 |
0.802 |
38 |
1.153 |
19 |
0.597 |
29 |
0.830 |
39 |
1.202 |
Table 2 Inspection Results of Saturated Water in Benzene for Different Sample Introduction Quantity
Serial Number |
A1 |
A2 |
A3 |
B1 |
B2 |
B3 |
C1 |
C2 |
C3 |
mass before sample injection (mg) |
|
|
|
|
|
|
|
|
|
mass after sample injection (mg) |
|
|
|
|
|
|
|
|
|
mass of sample injection (mg) |
|
|
|
|
|
|
|
|
|
temperature of sample injection (℃) |
|
|
|
|
|
|
|
|
|
theoretical concentration of water (μg/mg) |
|
|
|
|
|
|
|
|
|
average value of theoretical concentration of water(μg/mg) |
|
|
|
||||||
actual measured mass of water in benzene(μg) |
|
|
|
|
|
|
|
|
|
actual measured concentration of water in benzene (μg/mg) |
|
|
|
||||||
actual measured average value of water concentration (μg/mg) |
|
|
|
||||||
actual measured average value of water concentration(μg/mg)-theoretical concentration of water(μg/mg) |
|
|
|
||||||
basic error(%) |
|
|
|
Postulating the absolute water content of sampling is X, the relative error is
7. Measurement Method of Water Content in Sample
7.1 Reference Table for Sample Injection Quantity
Water Content |
Sample Injection quantity |
100% |
10mg |
50% |
20 mg~10 mg |
10% |
100 mg~10 mg |
1% |
|
0.1% |
|
0.01% |
|
0.001% |
|
0.0001% |
|
7.2 Measurement of Moisture in the Liquid Sample
7.2.1 Firstly, prepare intensive drying sample injection instrument with a pinhead(you can choose other sample injection instrument of different volume according to different determined sample), flush it 4~5 times with sample, then intake a certain volume of sample, prepared for sample injection.
7.2.2 Press ” SAMPLE INDUCTION” key.
7.2.3 Sample is injected to anode solution cell through sample injection instrument, be careful that needlepoint can not insert into the liquid(especially the sample that contains more water), after injecting sample, needlepoint can instantaneously contact liquid level to scour off residual sample of needlepoint, and avoid to contact the inner wall of titration cell or electrode, after injecting, the electroanalysis will automatically start when delay time ends, the ring of buzzer means the electroanalysis is finished.
7.2.4 The unit of measurement result is “microgramme water”, the calculation formula of concentration is as follows:
measurement result (ug) meassured result (ug) meassured result (ug)
water content(ppm) = ———————— = ——————————————————
mass of sample (g) sample density (g/mL)×sample volume (mL)
7.3 Measurement of Moisture in the Solid Sample
The shape of solid sample can be powder、granule、lump(big lump should be shattered), when sample is difficulty to dissolve in electrolytic solution, we should choose a appropriate water evaporator, and connect it to the main unit, and choose appropriate connector and sampling method according to species of sample.
Figure 4
7.3.
7.3.2 Taking out the cover of solid sample injection instrument,after sample is loaded, closing immediately,then accurate weighing the solid sample injection instrument with sample, recording numerical value.
7.3.3 Taking out plug of titration cell and cover of sample injection instrument,sample injection instrument is inserted into the entrance of sample according Figure 5.
7.3.4 After instrument is stabilized,press “SAMPLE INDUCTION” key, delay time begins,it is assured that solid sample can dissolve sufficiently during delay time.
7.3.5 Delay time begins, sample injection instrument is rotated 180 degree(as shown by broken line in Figure 5),sample is dropped into electrolytic solution, (when solid sample is dropped into electrolytic solution, solid sample can not contact wall of titration cell and cathode cell、measuring electrode). After delay ends, measurement begins.
Figure 5
7.3.6 After the unit is balanced, and reading number accurately, sample injection instrument is taken out carefully, and covering injection plug of sample, and accurate weighing solid sample injection instrument, reading number, the subtraction between this numerical value and weighing value in Step 2 is the mass of sample injection.
7.3.7 The measurement method of water content in solid is the same as the method of liquid.
7.4 Measurement method of water content in gas sample
When measuring water content in gas sample, 150mL electrolytic solution should be injected into the anode cell to make sure that the moisture in gas is sufficiently ingested. Simultaneity, flow speed of gas is controlled to 1
When measuring trace moisture in liquid hydrocarbon, it is suggested that you should use our company’s “sampling and sample injection instrument of liquid hydrocarbon flash distillation and gasification” simultaneously.
8 Attention
8.1 Attention for Electrolytic Solution
8.1.
8.1.2 For the electrolytic solution in cathode cell,if it is found that strong air bubble or polluted and light mahogany electrolyte is released in the course of titration, blank electric current will be increased, measurement reappearance is reduced, time of reaching end point is lengthened, electrolytic should be replaced in the instance.
8.1.3 If titration time exceeds half of an hour, the unit can not be steady, agitating is stopped by rotating agitating button,observe if brown iodine is generated on anode platinum net, if there is no iodine or few, electrolytic solution should be replaced.
8.1.4 Be careful that do not inhaling electrolytic solution or contact it by hand, if contacting it with skin, it is flushed thoroughly by water, because there is poisonous component in the reagent, the laboratory should be well draughty.
8.2 Attention for Measurement
8.2.1 When sample is injected into titration cell, needlepoint of sample injection instrument can not be inserted to liquid level(especially sample in which water content is bigger), After sample is injected, needlepoint can contact liquid level momentary, residual sample on the needlepoint is flushed. When liquid, solid and gas are injected, sample should be contacted to inner wall of titration cell and electrode.
8.2.2 The typical measurement range of the unit is 3ug-100mg, In order to obtain accurate measurement result, you should control injection volume of sample according to water content of sample.
8.2.3 Cleaning out residual liquid on the silica gel pad of sample injection port in time.
9 Maintenance and Inspection
9.1 Maintenance of Electrolytic Solution
The electrolyte is located in the draughty place, the temperature should under
9.2 Replacement of Silica-gel Pad
The silica-gel pad of sample injection port is used for a long time, the pinhole of it will lose contractibility, thus, the moisture in atmosphere will go into titration cell, which can lead to measurement error, the pad should be replaced
9.3 Replacement of Silica-gel
When color of silica gel in drying pipe changes from blue to light blue or red, the silica-gel should be replaced
9.4 Maintenance for the Abrasive Mouth of Titration Cell
Turning the connect of abrasive mouth of titration cell in a week, when it can not be turned easily, it is disassembled carefully, clean the abrasive mouth and faucet, and thin vacuum grease is smeared again(attention: you should not smear too much vacuum grease, otherwise it enters into titration cell and causes measurement error), if you do not check in time, vacuum grease becomes rigidity after a long time, the components connected by abrasive mouth will agglutinate each other and can not be disassembled. So it should be maintained well, and it is easy to be disassembled and be cleaned.
9.5 Treatment of Agglutination of Titration Cell’s Abrasive Mouth
If the abrasive mouths of titration cell agglutinates firmly, it can not be disassembled, you can follow the following procedures to solve this problem:
9.5.1 Draining off electrolytic of titration cell, and cleaning up it.
9.5.2 Injecting a few acetone around abrasive mouth, then turning it lightly by hand, thus, it can be disassembled.
9.5.3 If it can not be disassembled, you put titration cell into a
9.6 Maintenance of Detector electrode
leakage |
short circuit |
short circuit |
easily broken
position |
9.6.1 When magnetic force stirrer rotates in high speed, you should pay attention that mixing propeller can be bounced to damage electrode.
9.6.2 When detector electrode is entered into or taken out, power source should be turned off, and pay attention that not contacting detector electrode to inner wall of titration cell.
9.6.3 When detector electrode is bent but not short circuit, it can be used or repaired. When repairing, gripping root of platinum electrode with forceps, slowly collating top end of platinum electrode.
9.6.4 When detector electrode is polluted, wiping detector electrode with acetone, if dirt on platinum thread can not be removed, burning end of platinum thread with flame of alcohol burner. (attention: flame is approach to end of platinum thread slowly, to avoid breaking of electrode glass because of heating up rapidly).
9.7 Maintenance of Cathode Cell
8.7.1 Disassembling cathode cell
When disassembling cathode cell, you should be slowly and should not touch the top end and inner wall of titration, because platinum thread and platinum net are stretched from the interface of cathode cell’s abrasive mouth.
9.7.2 Cleaning of Cathode Cell
When cathode cell is polluted, including the following cases:
(1)reducing efficiency of electrolysis, delaying measurement time.
(2)blank electric current is increased because polluted part adsorbs moisture,.
(3)titration speed is not steady, and not arriving end point.
If above-mentioned cases occurs, cleaning dirt of glass part’s surface and platinum net with chloroform(attention: do not bump platinum thread and platinum net), pouring chloroform into cathode cell, sealing interface of drying pump with rubber plug or similarity stuff, swaying sufficiently to remove inner dirt(this can be repeatedly). Then pouring chloroform onto the whole surface of glass part to clean, but not cleaning to down-lead of electrode. When it still cannot be cleaned, you can immerge cathode cell into beaker with diluted sulfuric acid (pay attention not to contacting platinum thread and platinum net).
position difficult
to be dried |
easily broken
position |
top
end |
platinum
thread |
platinum net |
pore wall |
9.7.3 Drying of Cathode Cell
Drying cathode with hot wind of air blower, the moisture in sand core is difficulty to dry, so it should be dried thoroughly. When there is residual moisture, put cathode cell into vaccum desiccator to desiccate 12 hours.
9.8 Maintenance of Electrode Plug and Socket
The plug and socket of detector electrode and cathode cell electrode is used frequently, this causes the loosening of plug gradually. The dust will adhere to the plug and socket after a long time use, it makes that plug and socket of electrode is bad contact, so it should be cleaned.
9.8.1 Loosening of Plug
When connection of plug and socket is loose, you can press outer sheet metal of plug inwardly with vice.
9.8.2 Cleaning of Plug and Socket
Wiping metal part with ethanol or acetone, which can make a good contact.
sheet metal |
10 Troubleshooting
10.1 Open Circuit of Measurement and Titration
When open circuit of measurement and titration begins,state shows it will increase rapidly to maximum,“electrode open circuit” is showed in display window.
10.1.1 For this case, you should check whether the contact between plug of detector electrode and cathode cell are normal.
10.1.2 Checking that down-lead of detector electrode and electrode of cathode cell are good contact or bad contact.
10.2 Measurement of Short Circuit
When measuring short circuit, short circuit of electrode is showed in the unit.
10.2.1 Checking whether detector electrode of plug and socket is short circuit.
10.2.2 Checking whether electrode is short circuit.
10.2.3 Checking whether electrode is leak which causes that the unit can not arrive end point. If detector electrode is leak, despite titration time excess half an hour, the unit can not arrive end point too(this case does not belong to reagent problem).
10.2.4 When new electrolyte is injected, short circuit of electrode may appear, for this problem, you can slowly drop a little distilled water until show normally.
10.3 Electroanalysis of Open Circuit
When electrolyzing open circuit, open circuit of electrolyzing is showed in the unit.
10.3.1 Checking whether Electrolysis electrode of plug and socket is connected.
10.3.2 Checking that down-lead of Electrolysis electrode and electrode of anode cell are good contact or bad contact.
10.4 For other malfunctions, contact our company.
Copyright @ 2013 Daqing City Rishang Instrument Manufacturing Co., Ltd. All rights reserved Power:兴旺宝明通